Magnetrol 2xx Specifications Page 35

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 34
46-648 Jupiter Magnetostrictive Transmitters
35
2.7.2 Functional
Measured variables Continuous liquid level
Input power (at terminals) 12-28 VDC
Signal output 4–20 mA
4–20 mA with HART 5.0
NAMUR NE 43 compliant with 3.8 to 21.5 mA useable range
Loop resistance 620 maximum ohms @ 24 VDC—refer to chart below
Power consumption 0.7 watt, refer to chart below
Damping 0 to 25 seconds
Error signal 3.6 or 22 mA, field selectable
User interface 3-button keypad, HART communicator, AMS software, PACTware
or FOUNDATION Fieldbus
Display 2-line × 8-character LCD in inches or cm, mA, and % of level
Resolution Analog: 0.01 mA
Digital: 0.01 units
Span 6 to 400 inches (999 cm)
SIL 2 Safe Failure Fraction (SFF) 90.7%
(consult factory for SIL safety manual)
1200
1000
800
600
400
200
0
0 10 20 30 40 VDC
20.5 mA
24 VDC
620
11
Power Consumption
See Bulletin 46-649 for
F
OUNDATION Fieldbus output
Enclosure type Dual compartment
Enclosure material Sand cast aluminum grade 356 HT or 316 stainless steel
Enclosure finish Baked on polymer powder coat
Enclosure rating NEMA 4X7/9, IP 66
Sensor material 316 stainless steel
Sensor length 6 to 400 inches (15 to 999 centimeters)
2.7.3 Physical