Magnetrol 2xx Specifications Page 29

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 28
46-648 Jupiter Magnetostrictive Transmitters
29
Unit with Display
Communicates
Error
Displayed?
No Level Signal
Error Codes
No NoYes
Reload all
Parameters
Yes
Yes Yes
Corrupt Display
Text?
Replace Electronic
Module
No
Yes
Replace Electronic
Module
Same Error
Message?
Retest No
Yes
Out of Calib
Analog Loop
Correct?
Trim 4-20 mA
No
Yes
No
Call Factory
Display Reading
Correctly?
4
Yes
No
Unit with Display
Communicates
3
Check probe
length
Float is not
detected
Consult Factory
Check Offset
and
4-20 mA
Set Points
Page view 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 39 40

Comments to this Manuals

No comments